A molecular laboratory can detect a few copies of a target, which makes it vulnerable to tiny amounts of contamination. Good design controls where material is allowed to go.
The core principle: one direction
Reagents and clean consumables should move towards samples, never the other way, and people and equipment should not carry material back from amplification areas to clean ones. This is achieved by physical separation, dedicated equipment and lab coats, and clear signage.
Typical zones
- Reagent preparation: the cleanest area, where master mixes are prepared.
- Sample preparation: sample receipt, inactivation and nucleic acid extraction, with biosafety controls.
- Amplification and detection: where PCR runs. With closed tubes or cartridges, amplified product stays contained.
- Post-amplification: only if the workflow involves open handling of amplified product, for example some sequencing preparation.
How cartridge systems change things
Integrated cartridge systems such as the Ultra Multiplex PCR keep extraction, amplification and detection inside a closed cartridge. This reduces open handling and can simplify layouts for smaller hospitals and mobile units, although good practice for sample handling and waste still applies.
Other design inputs
- Power with backup for instruments and freezers
- Air handling and temperature control
- Workflow for samples, people and waste
- Space for storage, documentation and growth
Our turnkey lab design service covers all of this, from layout to commissioning.